Máy dò tìm rò rỉ (Máy kiểm tra rò rỉ) Helium Leak Detector LINXON LX218
Giá bán: Liên Hệ.
- Mô tả sản phẩm
Helium Leak Detector LINXON® LX218
Robust high-performance leak detector at a fair price
Unboxing Helium Leak Detector LX218
Installation Helium Leak Detector LX218
LINXON LX218 – advantages at a glance
Integrated INFICON components enable first-class performance for absolute measurement accuracy at outstanding speed. The simple navigation and robust construction make the LX218 leak detector the ideal choice for companies in a wide range of industries. The helium leak detector has a start-up time of less than three minutes and offers a fast leak test and reaction time. A further advantage: With its vacuum mode and sniffing option, it is perfectly suited for efficient leak localization and leak size determination.
- Competitive price thanks to standard components and direct distribution channels
- Highest degree of availability and reliability, even after long periods of use, with proven vacuum technology components such as mass spectrometers, pumps and valves
- Very short measuring times due to high pumping speed
- Detection of the smallest leaks, even in environments contaminated by helium, with an intelligent stainless steel vacuum system
- The highest quality in your production or service application guaranteed by high-precision INFICON measuring technology
- Easy operation and installation, with no training required
- Robust and versatile, it can be used even in the harshest production environments
ORDER INFORMATION BASIC UNIT | |
LINXON LX218 WET 230V 50Hz version 115V 60Hz version 100V 60Hz version LINXON LX218 DRY 230V 50Hz version 115V 60Hz version LINXON LX218 MODULE 100V-250V 50/60 Hz | 8200-000 8200-001 8200-005
8200-002 8200-003
8200-004 |
TECHNICAL DATA | |
Lowest detectable leak rate for He (according to AVS 2.1) | ‹ 5 x 10-13 Pa m3/s |
Ion source | Two long-life iridium filaments, yttrium-oxide coated (three-year warranty) |
Testing method | Vacuum and sniffer leak detection |
Detectable gases | 4He, 3He, H2 |
Internal He test leak | 10-8 Pa m3/s |
Leak rate display | 10-13 – 0.1 Pa m3/s |
Time to operation | 3 min. |
Response time | 0.5 s |
Suction capacity: He | 2.5 l/s |
Connecting flange | DN 25 ISO-KF |
Inlet pressure, max. | 25 mbar |
Ambient temperature | +10°C to +35°C in operation |
Input: analog | Compact gauge |
Input: digital | Start/stop, zero, calibration |
Interface | RS-232-C, RS-485 |
Output: analog | Leak rate signal 2 x 0-10 V, lin/log |
Output: digital | Ready to start, ready to measure, leak, error, partial flow valve |
Relay: switch voltage | 2 x 230 V AC |
Relay: switch current | 3 A |
ORDER INFORMATION ACCESSOIRES | |
Sniffer line 3m | 551-300 |
Sniffer line 5m | 551-301 |
Sniffer line 10m | 551-302 |
Remote control with cable (INFICON RC1000) | 551-010 |
Remote control wireless (INFICON RC1000) | 551-015 |