Bộ đo áp suất chân không Vacuum Pressure Gauges INFICON Cold Cathode Gauges MAG470
Giá bán: Liên Hệ.
- Mô tả sản phẩm
Vacuum Pressure Gauges INFICON Cold Cathode Gauges MAG470
“Inverted Magnetron Gauges”
The Inverted Magnetron Vacuum Gauge MAG470 is for all ultra high vacuum applications. The measurement range is from 5×10-11 mbar to 1×10-2 mbar.
Advantages
- No filament to burn out
- Extended version, sensor bakeout up to 250°C
- Excellent ignition properties
- Easy to clean
- Metal-sealed feedthrough
- LED indicator for high voltage on
- Logarithmic analog output signal
Typical applications
- Ultra high vacuum pressure monitoring
- Base pressure for evaporation and sputtering systems
- General vacuum measurement and control in the medium and high vacuum range
Specifications
MAG470
Measurement range (air, N2) | mbar Torr | 5×10-11 … 1×10-2 3.8×10-11 … 7.5×10-3 |
Accuracy | ||
1×10-9 … 1×10-3 mbar | % of reading | ≈±30 |
Repeatability | ||
1×10-9 … 1×10-3 mbar | % of reading | ≈±5 |
Temperature | ||
Operation (ambient) | °C | +5 … +55 |
Storage | °C | -40 … +65 |
Bake-out | °C | 250 1) |
Supply voltage | ||
At gauge | V (dc) | +15 … +30 |
At supply unit with max. cable length (3) | V (dc) | +16 … +30 |
Ripple | V (p-p) | ≤1 |
Power consumption | W | ≤2 |
Materials exposed to vacuum | Stainless steel, Al2O3, Ag | |
(1) Without electronics magnetic shielding (2) 250°C in defined area, see dimension (3) The minimum voltage of the supply unit must be increased proportionally to the length of the sensor cable |
MAG470 extended
Measurement range (air, N2) | mbar Torr | 5×10-11 … 1×10-2 3.8×10-11 … 7.5×10-3 |
Accuracy | ||
1×10-9 … 1×10-3 mbar | % of reading | ≈±30 |
Repeatability | ||
1×10-9 … 1×10-3 mbar | % of reading | ≈±5 |
Temperature | ||
Operation (ambient) | °C | +5 … +55 |
Storage | °C | -40 … +65 |
Bake-out | °C | 250 2) |
Supply voltage | ||
At gauge | V (dc) | +15 … +30 |
At supply unit with max. cable length (3) | V (dc) | +16 … +30 |
Ripple | V (p-p) | ≤1 |
Power consumption | W | ≤2 |
Materials exposed to vacuum | Stainless steel, Al2O3, Ag | |
(1) Without electronics magnetic shielding (2) 250°C in defined area, see dimension (3) The minimum voltage of the supply unit must be increased proportionally to the length of the sensor cable |
Part Number
Part Number
MAG470 | |
Part Number | Description |
399-360 | MAG470 DN 40 CF-F, FCC |
MAG470 extended | |
Part Number | Description |
399-361 | MAG470 DN 40 CF-F, extendet |
Accessories
MAG470 | |
Part Number | Description |
351-023 | Magnetic shield |
MAG470 extended | |
Part Number | Description |
351-023 | Magnetic shield |
Spare Parts
MAG470 | |
Part Number | Description |
351-997 | MPG401 Maintenance set |
351-990 | Repair set IKR 060/070/26x/270 MAG470 |
351-995 | MxG40x/50x Spare parts ignition aid (10) |
351-994 | MPG40x Ignition tool set |
MAG470 extended | |
Part Number | Description |
351-990 | Repair set IKR 060/070/26x/270 MAG470 |
351-995 | MxG40x/50x Spare parts ignition aid (10) |
351-994 | MPG40x Ignition tool set |
351-997 | MPG401 Maintenance set |